Published September 1987 | Version v1
Journal article

Shielding requirements for mammography

Creators

  • 1. St. Luke's Hospital, Milwaukee, WI

Description

Shielding requirements for mammography installations have been investigated. To apply the methodologies of NCRP Report No. 49, the scatter-to-incident ratio of a typical mammography beam was measured, and the broad beam transmission was calculated for several representative beam spectra. These calculations were found to compare favorably with published low kVp tungsten-targeted x-ray transmission through a variety of shielding materials. Radiation shielding tables were developed from the calculated transmissions through Pb, concrete, gypsum, steel, plate glass, and water, using a technique which eliminates the ''add one HVL'' rule. It is concluded that Mo-targeted x-ray beams operated at 35 kVp require half the shielding of W-targeted beams operated at 50 kVp, and that adequate, cost-effective shielding calculations will consider alternatives to Pb

Additional details

Publishing Information

Journal Title
Health Phys.
Journal Volume
53
Journal Issue
3
Series
Health Phys.
Journal Page Range
267-279
ISSN
0017-9078
CODEN
HLTPA