Shielding requirements for mammography
Description
Shielding requirements for mammography installations have been investigated. To apply the methodologies of NCRP Report No. 49, the scatter-to-incident ratio of a typical mammography beam was measured, and the broad beam transmission was calculated for several representative beam spectra. These calculations were found to compare favorably with published low kVp tungsten-targeted x-ray transmission through a variety of shielding materials. Radiation shielding tables were developed from the calculated transmissions through Pb, concrete, gypsum, steel, plate glass, and water, using a technique which eliminates the ''add one HVL'' rule. It is concluded that Mo-targeted x-ray beams operated at 35 kVp require half the shielding of W-targeted beams operated at 50 kVp, and that adequate, cost-effective shielding calculations will consider alternatives to Pb
Additional details
Publishing Information
- Journal Title
- Health Phys.
- Journal Volume
- 53
- Journal Issue
- 3
- Series
- Health Phys.
- Journal Page Range
- 267-279
- ISSN
- 0017-9078
- CODEN
- HLTPA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 19017490
- Subject category
- S63: RADIATION, THERMAL, AND OTHER ENVIRONMENTAL POLLUTANT EFFECTS ON LIVING ORGANISMS AND BIOLOGICAL MATERIALS; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; GLASS; LEAD; MAMMARY GLANDS; RADIATION DOSES; RADIATION PROTECTION; SHIELDING; WATER; X RADIATION
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; ELEMENTS; GLANDS; HYDROGEN COMPOUNDS; IONIZING RADIATIONS; MEDICINE; METALS; ORGANS; OXYGEN COMPOUNDS; POLAR SOLVENTS; RADIATIONS; SOLVENTS