Published May 2017 | Version v1
Journal article

Tetrazolium salt monomers for gel dosimetry I: Principles

  • 1. Department of Chemical and Biochemical Engineering, University of Western Ontario, London, ON, N5A 5B9 (Canada)

Description

Tetrazolium salts (TS) have been previously used for radiochromic dosimetry in solutions, films and three dimensional (3D) gelatine-based gels. However, widespread application for 3D dosimetry has not been achieved due to the required high concentrations and associated high costs of the TS dimer used in prior research. Through careful selection of TS monomer, sensitivity-enhancing additives and inert gel forming material, we report the preparation of a non-diffusing, chemically stable, 3D dosimeter with linear sensitivity between 0 and 80 Gy with submillimolar requirements for the active TS. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/847/1/012048

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
847
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1742-6596

Conference

Title
9. international conference on 3D radiation dosimetry
Dates
7-10 Nov 2016
Place
Galveston, TX (United States)