Feasibility of prospectively ECG-triggered high-pitch coronary CT angiography with 30 mL iodinated contrast agent at 70 kVp: initial experience
Creators
- 1. Medical School of Nanjing University, Department of Medical Imaging, Jinling Hospital, Nanjing, Jiangsu (China)
- 2. Medical School of Nanjing University, Department of Cardiology, Jinling Hospital, Nanjing, Jiangsu (China)
- 3. Medical University of South Carolina, Department of Radiology and Radiological Science, Charleston, SC (United States)
Description
To evaluate the feasibility, image quality and radiation dose of prospectively ECG-triggered high-pitch coronary CT angiography (CCTA) with 30 mL contrast agent at 70 kVp. Fifty-eight patients with suspected coronary artery disease, a body mass index (BMI) of less than 25 kg/m2, sinus rhythm and a heart rate (HR) of less than 70 beats per minute (bpm) were prospectively enrolled in this study. Thirty mL of 370 mg I/mL iodinated contrast agent was administrated at a flow rate of 5 mL/s. All patients underwent prospectively ECG-triggered high-pitch CCTA on a second-generation dual-source CT system at 70 kVp using automated tube current modulation. Fifty-six patients (96.6 %) had diagnostic CCTA images and two patients (3.4 %) had one vessel with poor image quality each rated as non-diagnostic. No significant effects of HR, HR variability and BMI on CCTA image quality were observed (all P > 0.05). Effective dose was 0.17 ± 0.02 mSv and the size-specific dose estimate was 1.03 ± 0.13 mGy. Prospectively ECG-triggered high-pitch CCTA at 70 kVp with 30 mL of contrast agent can provide diagnostic image quality at a radiation dose of less than 0.2 mSv in patients with a BMI of less than 25 kg/m2 and an HR of less than 70 bpm. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-014-3157-2Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 24
- Journal Issue
- 7
- Journal Page Range
- p. 1537-1546
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 45081657
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AUTOMATION; BLOOD FLOW; CALCIUM; COMPUTERIZED TOMOGRAPHY; CONTRAST MEDIA; CORONARIES; DIAGNOSIS; ELECTROCARDIOGRAMS; FEASIBILITY STUDIES; FLOW RATE; IODINATION; MILLI GY RANGE 10-100; PLAQUE FORMATION; RADIATION DOSES; VASCULAR DISEASES
- Descriptors DEC
- ABSORBED DOSE RANGE; ALKALINE EARTH METALS; ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR DISEASES; CARDIOVASCULAR SYSTEM; CHEMICAL REACTIONS; DIAGNOSTIC TECHNIQUES; DIAGRAMS; DISEASES; DOSES; ELEMENTS; HALOGENATION; INFORMATION; METALS; MILLI GY RANGE; ORGANS; RADIATION DOSE RANGES; TOMOGRAPHY