Published May 1982 | Version v1
Journal article

Sensitive large-area calorimeter for measuring the energy of laser-produced plasma soft X radiation

Description

A thermoelectric detector for measuring the energy of laser-produced plasma soft X radiation is described. The measurements are made in the wave length range, lambda=1.5 to 14 A. A small thickness of the detecting element (approximately 100 mcm Al), great number of copper-constantan thermocouples and large area (about 25 cm2) allow to obtain a high sensitivity of the instrument equal to 10-4 J/cm2. The absolute calibration of the instrument was performed in a vacuum chamber at a pressure < or approximately 0.1 torr by the substitution method. Total error when measuring the X radiation energy of about 10-4 J is 15%; for the energy of about 10-2 J it is 5%. The conversion coefficient of the laser radiation into the plasma X radiation for a teflon target has been determined to be about 3%

Additional details

Additional titles

Original title (Russian)
Чувствительный калориметр большой площади для измерения энергии мягкого рентгеновского излучения лазерной плазмы

Publishing Information

Journal Title
Prib. Tekh. Ehksp.
Journal Issue
no.3
Series
Prib. Tekh. Ehksp.
ISSN
0032-8162

Optional Information

Notes
For English translation see the journal Instruments and Experimental Techniques (USA).