Optimisation of dose distribution in single planar iridium implants
Description
Single planar interstitial implants aim to irradiate 10 mm thickness of tissue to uniform dose. Interstitial iridium implants are normally practised with the same linear activity throughout the implant. As a result, the peripheral regions of the implant get lesser dose contribution from the central wires. Our study with the planning system for a 5 x 4 cm2 area of parallel implant geometry shows that the peripheral regions receive only 83% of the prescribed dose and the minimum treated thickness is 7.6 mm only. Increasing either the activity or time weight to the peripheral wires by 25% enables the peripheral regions to receive a minimum of 92% of the prescribed dose and the minimum treated thickness increases to 8.80 mm. The removal time of peripheral wires with time weight option can be calculated manually by a simple formula. (author). 2 refs., 6 figs., 3 tabs
Additional details
Publishing Information
- Journal Title
- Journal of Medical Physics
- Journal Volume
- 21
- Journal Issue
- 1
- Journal Page Range
- p. 31-34.
- ISSN
- 0971-6203
- CODEN
- JMPHFE
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 28008309
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DOSE RATES; GEOMETRY; IRIDIUM; ISODOSE CURVES; RADIATION DOSE DISTRIBUTIONS; RADIATION SOURCE IMPLANTS; RADIOTHERAPY; THICKNESS
- Descriptors DEC
- DIMENSIONS; ELEMENTS; IMPLANTS; MATHEMATICS; MEDICINE; METALS; PLATINUM METALS; RADIATION SOURCES; THERAPY; TRANSITION ELEMENTS