Published August 2019 | Version v1
Journal article

Estimation of patient lens dose associated with C-arm cone-beam computed tomography usage during interventional neuroradiology

  • 1. Course of Radiological Technology, Health Sciences, Graduate School of Medicine, Tohoku University, 2-1 Seiryo, Aoba, Sendai, Miyagi 980-8575 (Japan)
  • 2. Department of Radiology, Toranomon Hospital, 2-2-2 Toranomon, Minato, Tokyo 105-8470 (Japan)
  • 3. Department of Radiological Health Science, Institute of Industrial Ecological Sciences, University of Occupational and Environmental Health, 1-1, Iseigaoka, Yahatanishi, Kitakyushu, Fukuoka 807-0804 (Japan)
  • 4. Division for Stroke Prevention and Treatment, Department of Neurosurgery, Faculty of Medicine, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8575 (Japan)
  • 5. Faculty of Medicine, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8575 (Japan)
  • 6. Department of Endovascular Neurosurgery, Toranomon Hospital, 2-2-2 Toranomon, Minato, Tokyo 105- 8470 (Japan)

Description

The purpose of this study was to investigate the dose distribution and lens doses associated with C-arm cone-beam computed tomography (CBCT), using a head phantom, and to estimate the contribution ratio of C-arm CBCT to each patient's lens dose during interventional neuroradiology ('lens dose ratio') in 109 clinical cases. In the phantom study, the peak skin doses and respective right and left lens doses of C-arm CBCT were as follows: 63.0 ± 1.9 mGy, 19.7 ± 1.4 mGy and 21.9 ± 0.8 mGy in whole brain C-arm CBCT and 39.2 ± 1.4 mGy, 4.7 ± 0.9 mGy and 3.6 ± 0.3 mGy in high-resolution C-arm CBCT. In the clinical study, the lens dose ratios were 25.4 ± 8.7% in the right lens and 19.1 ± 9.8% in the left lens. This study shows that, on average, ∼25% of patients' total lens dose was contributed by C-arm CBCT. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncy188

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
184
Journal Issue
2
Journal Page Range
p. 138-147
ISSN
0144-8420

Conference

Title
15. Radiation Emergency Medical Preparedness and Assistance Network (REMPAN) Meeting
Dates
3-5 Jul 2017
Place
Geneva (Switzerland)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
France
INIS RN
50066346
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; EYES; PATIENTS; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DOSES; FACE; HEAD; MOCKUP; ORGANS; SENSE ORGANS; STRUCTURAL MODELS; TOMOGRAPHY

Optional Information

Notes
28 refs.