Published 1989 | Version v1
Report Open

Elimination of the effect in a proportional neutron detector using surface charge on insulators

Description

The proportional neutron detector of 1500 cm3 volume filled with a gas mixture containing 3He is described. Inhomogeneities of electric field near the ends are avoided using the surface charge created by positive ions on teflon insulators covering the total surface of ends. The energy resolution of this detector is 2.8% for isotopic flux of thermal neutrons at the registration of 3He(n,α)T reaction products. 6 refs.; 3 figs

Availability note (English)

MF available from INIS under the Report Number.

Files

22061053.pdf

Files (141.8 kB)

Name Size Download all
md5:68200fd519ddc94df2a29731cb3760b7
141.8 kB Preview Download

System files (14.0 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Устранение торцового эффекта в пропорциональном счетчике нейтронов с помощью поверхностного заряда на изоляторах

Publishing Information

Imprint Pagination
4 p.
Report number