Published February 1988 | Version v1
Journal article

Imaging characteristics of a high resolution cone beam collimator

  • 1. P.O. Box 3949, Duke Univ. Medical Center, Durham, NC (USA)

Description

Single photon emission computed tomography (SPECT) using cone beam geometry requires accurately built collimators and an appropriate reconstruction algorithm. The authors have investigated the imaging characteristics of a cast, high resolution, cone beam collimator, and have compared its measured sensitivity and resolution with analytically predicted values. Using a filtered backprojection algorithm, the authors have measured point spread functions at several locations within the field-of-view. Measurements have been made both with the collimator surface parallel to the axis-of-rotation (AOR) and also with the surface tilted at an angle of 20 degrees with respect to the AOR. Although the SPECT point spread response is not completely shift-invariant, high quality cone beam images of a brain phantom have been obtained

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
35
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
644-648
ISSN
0018-9499
CODEN
IETNA

Conference

Title
34. nuclear science symposium and 19. nuclear power systems symposium.
Dates
21-23 Oct 1987.
Place
San Francisco, CA (USA).

Optional Information

Secondary number(s)
CONF-871006--.