Description
In this paper, we present the design of a novel low-noise front-end readout application-specific integrated circuit (ASIC) for our small animal PET systems which objective is to achieve the following performances, the spatial resolution of 1 mm3, the detection efficiency of 15 % and the time resolution of 1 ns. A cascade amplifier based on the PMOS input transistor is selected to realize the charge-sensitive amplifier (CSA) for the sake of good noise performances. The output of the CSA is split into two branches. One is connected to a slow shaper for energy measurements. The other is connected to a fast shaper for time acquisition. A novel monostable circuit is designed to adjust the time delay of the trigger signals so that the peak value of the shaped voltages can be sampled and stored. Based on the above techniques, an eight-channel front-end readout prototype chip is designed and implemented in 0.35 μm CMOS process. The die size is 2.286 mm x 2.282 mm. The input range of the ASIC is from 2000 e- to 180000 e-, reflecting to the energy of the gamma ray from 11.2 keV to 1 MeV. The gain of the readout channel is 65 V/pC. The best test result of ENC is 86.5 e- at zero farad plus 9.3 e- per pico-farad. The nonlinearity is less than 3 %. The crosstalk is less than 2 %. The power dissipation is about 9 mW/channel (authors)
Files
Additional details
Identifiers
Publishing Information
- Imprint Pagination
- 6 p.
- Report number
- ANIMMA--2013-1150
Conference
- Title
- 3. international conference on Advancements in Nuclear Instrumentation Measurement Methods and their Applications
- Acronym
- ANIMMA 2013
- Dates
- 23-27 Jun 2013
- Place
- Marseille (France)
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 46071519
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AMPLIFIERS; DESIGN; EFFICIENCY; GAMMA RADIATION; INTEGRATED CIRCUITS; KEV RANGE; MEV RANGE 01-10; NOISE; RADIATION DETECTION; READOUT SYSTEMS; SPATIAL RESOLUTION; TIME RESOLUTION; VANADIUM 65
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; DETECTION; ELECTROMAGNETIC RADIATION; ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; ENERGY RANGE; EQUIPMENT; INTERMEDIATE MASS NUCLEI; IONIZING RADIATIONS; ISOTOPES; MEV RANGE; MICROELECTRONIC CIRCUITS; MILLISECONDS LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; RADIATIONS; RADIOISOTOPES; RESOLUTION; TIMING PROPERTIES; VANADIUM ISOTOPES
Optional Information
- Notes
- 15 Refs.; Available from the INIS Liaison Officer for France, see the 'INIS contacts' section of the INIS website for current contact and E-mail addresses: http://www.iaea.org/inis/Contacts/