Published April 15, 1979 | Version v1
Journal article

Relative γ-ray intensities for calibration of Ge(Li) detectors

Creators

  • 1. Ceskoslovenska Akademie Ved, Rez. Ustav Jaderne Fyziky

Description

Mean relative γ-ray intensities for some widely used calibration standards were calculated from the various data published in the past ten years. The mutual consistency of the data under investigation was also analyzed. The mean γ-ray intensities and other relevant data are presented for the isotopes 56Co, 75Se, sup(110m)Ag, 133Ba, 152Eu, 160Tb, sup(166m)Ho, 169Yb, 182Ta, 192Ir, and 226Ra. The present mean γ-ray intensities were used for calibration of a 27 cm3 close-ended coaxial detector. (Auth.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods
Journal Volume
161
Journal Issue
1
Series
Nucl. Instrum. Methods.
Journal Page Range
151-159
ISSN
0029-554X