Published February 11, 1999 | Version v1
Journal article

Soil sample moisture content as a function of time during oven drying for gamma-ray spectroscopic measurements

Description

In routine gamma-ray spectroscopic analysis of collected soil samples, procedure often calls to remove soil moisture by oven drying overnight at a temperature of 100 deg. C . Oven drying not only minimizes the gamma-ray self-attenuation of soil samples due to the absence of water during the gamma-ray spectroscopic analysis, but also allows for a straightforward calculation of the specific activity of radionuclides in soil, historically based on the sample dry weight. Because radon exhalation is strongly dependent on moisture , knowledge of the oven-drying time dependence of the soil moisture content, combined with radon exhalation measurements during oven drying and at room temperature for varying soil moisture contents, would allow conclusions to be made on how the oven-drying radon exhalation rate depends on soil moisture content. Determinations of the oven-drying radon exhalation from soil samples allow corrections to be made for the immediate laboratory gamma-ray spectroscopy of radionuclides in the natural uranium decay chain. This paper presents the results of soil moisture content measurements during oven drying and suggests useful empirical fits to the moisture data

Additional details

Identifiers

PII
S0168900298010043;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
422
Journal Issue
1-3
Journal Page Range
p. 817-819
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 1998 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.