Effect of exposure factors on image quality in screening mammography
Creators
Description
Introduction: The aim of this research was to study the effect of exposure factors on image quality for digital screening mammography units in Kuwait which use Tungsten (W) targets with Rhodium (Rh) and Silver (Ag) as filters. Methods: Mammography Accreditation Phantom Model 015 was imaged using a Hologic Selenia Digital mammography unit with W targets and Rh and Ag filters. Four images, each at 26, 28, 30, and 32 kVp, were obtained using each target-filter combination (W/Rh and W/Ag). The images were evaluated by five senior technologists for the number of specks, fibers and masses visible on each image. Statistical analysis was carried out using non-parametric tests at p = 0.05 level. Results: There were significant changes in the visibility of fibers and specks between different kVp values with W/Rh (p < 0.001). However, with W/Ag combination, significant differences were observed in the fibers only (p < 0.001). Among the kVp values used, 28 kV emerged as the optimal value. Comparison of images obtained with the two filter materials, led to significant differences in the visibility of fibers and specks (p < 0.008). At 32 kVp, there were significant differences in the visibility of specks only (p < 0.008). Conclusion: A W/Rh target-filter combination provides better image quality than that provided by W/Ag. In particular, 30 and 32 kVp X-ray beams produce higher quality images than the lower kV values. - Highlights: • Visibility of specks improved significantly with W/Rh at 30 and 32 kVp compared to W/Ag. • A significant increase in the visibility of masses was observed when W/Rh is used with 32 kVp • A W/Rh target-filter combination provides better image quality than that provided by W/Ag.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radi.2017.05.005Additional details
Identifiers
- DOI
- 10.1016/j.radi.2017.05.005;
- PII
- S1078-8174(17)30062-7;
Publishing Information
- Journal Title
- Radiography (London 1995)
- Journal Volume
- 23
- Journal Issue
- 4
- Journal Page Range
- p. e99-e102
- ISSN
- 1078-8174
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 49049819
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; FIBERS; FILTERS; MAMMARY GLANDS; SCREENING; SILVER; SOCIO-ECONOMIC FACTORS; VISIBILITY; X RADIATION
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; ELEMENTS; GLANDS; INSTITUTIONAL FACTORS; IONIZING RADIATIONS; MEDICINE; METALS; NUCLEAR MEDICINE; ORGANS; RADIATIONS; RADIOLOGY; TRANSITION ELEMENTS
Optional Information
- Copyright
- Copyright (c) 2017 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.