Narrow photon beam dosimetry for linear accelerator radiosurgery

Description

The dosimetric characteristics of linear accelerator radiosurgery for 10-MV X-ray were measured. Measurement of the relative output factor and tissue maximum ratio with a microchamber produced results equivalent to those of measurement with X-ray film. The 80% isodose level width measured with the microchamber was significantly smaller than that measured with the X-ray film. For the measurement of relative output factor and tissue maximum ratio, a microchamber seems to be the more appropriate choice. X-Ray film was found to be suitable for beam profile measurement

Additional details

Identifiers

PII
S0167814096018439;

Publishing Information

Journal Title
Radiotherapy and Oncology
Journal Volume
41
Journal Issue
3
Journal Page Range
p. 221-224
ISSN
0167-8140
CODEN
RAONDT

Optional Information

Copyright
Copyright (c) 1996 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.