A new high sensitivity spectrometer for 3-D thermoluminescence analysis

  • 1. Milan Univ. (Italy). Ist. di Fisica
  • 2. AeDi Strumenti Scientifici, Lainate, MI (Italy)

Description

The features of a new spectrometer for measuring wavelength-resolved TL emission are described: the system is based on wide angle mirror optics and the dispersive element is a flat-field holographic grating, with numerical aperture f/2. The detector is a two-stage MCP (micro-channel plate) followed by a 512 photodiode array. The whole system, cooled by a Peltier control unit, reaches a signal-to-noise ratio comparable with a photomultiplier one. The high sensitivity of the apparatus allows the measurement in dosimetric materials, TL due to irradiations lower than 100 μGy. In quartz extracted from ancient ceramics the TL produced by natural irradiation has been detected. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
66
Journal Issue
1-4
Journal Page Range
p. 447-450.
ISSN
0144-8420
CODEN
RPDODE

Conference

Title
11. international conference on solid state dosimetry.
Dates
10-14 Jul 1995.
Place
Budapest (Hungary).

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
28034765
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
MIRRORS; OPTICAL SYSTEMS; READOUT SYSTEMS; SENSITIVITY; THERMOLUMINESCENT DOSEMETERS
Descriptors DEC
DOSEMETERS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS