Published 1996
| Version v1
Journal article
A new high sensitivity spectrometer for 3-D thermoluminescence analysis
Creators
- 1. Milan Univ. (Italy). Ist. di Fisica
- 2. AeDi Strumenti Scientifici, Lainate, MI (Italy)
Description
The features of a new spectrometer for measuring wavelength-resolved TL emission are described: the system is based on wide angle mirror optics and the dispersive element is a flat-field holographic grating, with numerical aperture f/2. The detector is a two-stage MCP (micro-channel plate) followed by a 512 photodiode array. The whole system, cooled by a Peltier control unit, reaches a signal-to-noise ratio comparable with a photomultiplier one. The high sensitivity of the apparatus allows the measurement in dosimetric materials, TL due to irradiations lower than 100 μGy. In quartz extracted from ancient ceramics the TL produced by natural irradiation has been detected. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 66
- Journal Issue
- 1-4
- Journal Page Range
- p. 447-450.
- ISSN
- 0144-8420
- CODEN
- RPDODE
Conference
- Title
- 11. international conference on solid state dosimetry.
- Dates
- 10-14 Jul 1995.
- Place
- Budapest (Hungary).
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 28034765
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- MIRRORS; OPTICAL SYSTEMS; READOUT SYSTEMS; SENSITIVITY; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- DOSEMETERS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS
Optional Information
- Secondary number(s)
- CONF--950708.