Published February 2019 | Version v1
Journal article

Analysis of failure of helical tomotherapy multi-leaf collimator system

  • 1. Department of Radiation Oncology, Peking Union Medical Hospital, Chinese Academy of Medical Sciences, Peking Union MeScal College, Beijing (China)

Description

Objective: To analyze the failure and processing methods of helical tomotherapy multileaf collimator (MLC) and summarize relevant maintenance experience, aiming to accurately identify the faulty parts, shorten the downtime and enhance the work time. Methods: The failure data of the helical tomotherapy MLC system in the past 48 months were analyzed to identify the common faulty parts, causes and processing methods. Results: During the previous 48 months, the MLC failure occurred for 20 times, 11 times for air compressor failure, 4 times for position verification board failure, twice for leaf driver failure, twice for cushion valve failure and once for the slip of leaf position verification rod. The MLC failure was significantly correlated with the humidity of high-pressure gas. The work time of machine exerted significant effect upon the service time of MLC parts. Conclusions: The structure of the helical tomotherapy MLC system is complex. The high-intensity work increases the failure rate. The humidity of high-pressure gas affects the normal operation of the MLC equipment. The faulty parts can be identified, the downtime can be reduced and the work time can be enhanced by summarizing the experience of MLC maintenance. (authors)

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiation Oncology
Journal Volume
28
Journal Issue
2
Journal Page Range
p. 155-157
ISSN
1004-4221

INIS

Country of Publication
China
Country of Input or Organization
China
INIS RN
55022172
Subject category
S07: ISOTOPES AND RADIATION SOURCES; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COLLIMATORS; CT-GUIDED RADIOTHERAPY; FAILURES; LINEAR ACCELERATORS
Descriptors DEC
ACCELERATORS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; RADIOTHERAPY; THERAPY

Optional Information

Notes
1 fig., 6 refs.; http://dx.doi.org/10.3760/cma.j.issn.1004-4221.2019.02.015