Published June 2002 | Version v1
Miscellaneous Open

Comparison of the air-kerma standards of the PTB and the BIPM in the medium-energy X-ray range

  • 1. Bureau International des Poids et Mesures (BIPM), 92 - Sevres (France)
  • 2. Physikakisch-Technische Bundesanstalt, Braunschweig (Germany)

Description

An indirect comparison has been made between the air erma standards of the PTB and the BIPM in the medium-energy x-ray range. The results show the standards to be in general agreement at the level of the stated standard uncertainty, although the result for the 100 kV radiation quality differs significantly from that for the other qualities. (authors)

Availability note (English)

Available from INIS in electronic form

Files

38101979.pdf

Files (226.8 kB)

Name Size Download all
md5:85fa563301f4e324e3ac5522518a9406
226.8 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
14 p.
Report number
BIPM--02-07

Optional Information

Notes
6 refs.