Published August 2013 | Version v1
Journal article

The MC simulations for SGS collimator best size

  • 1. Tile College of Nuclear Technology and Automation Engineering, Chengdu University of Technology, Chengdu (China)
  • 2. State Key Laboratory of Geohazard Prevention and Geoenvironment Protection, Chengdu University of Technology, Chengdu (China)

Description

In this paper, through the analysis of sub-scale theory of detection efficiency of the gamma scanning device (SGS), calculated the single photon detection efficiency formula, and analyzed the front plate collimator using MCNP5 software detection devices model calculations, simulated waste barrel in a different position, and tabled different spacing of the front plate collimator for high-purity germanium detector count rate, the simulation results show that: suitable for low-density material, flat collimator spacing is 3.5 cm; for medium-density substances, collimator suitable spacing is 4 cm; suitable for higher density material flat collimator spacing is 4.5 cm. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
33
Journal Issue
8
Journal Page Range
p. 1036-1038, 1042
ISSN
0258-0934