Published March 1986 | Version v1
Report Open

Test unit for measurement of small masses of alpha wastes

Description

New standards for waste surface storage require permanent management of fissile materials, even minute amounts, and accurate measurement of waste production. An accurate original system using gamma spectrometry is described in this article. It includes a lead shielded cell, a INa collimated sensor with direct counting including the peaks of Pu 239 and Pu 241 in the 400 keV range and data processing and recording. For a light matrix (cellulose or vinyl) sensitivity is 20 mg and accuracy +- 25 % and for a heavy matrix (graphite) 30 mg +- 40 %

Availability note (English)

MF available from INIS under the Report Number.

Abstract (French)

Compte tenu des nouvelles normes imposees pour le stockage des dechets α en surface il est necessaire de gerer en permanence les matieres fissiles, si petites soient elles et de mesurer avec precision les quantites de dechets des leur production. Le systeme presente est original au niveau de sa conception, il utilise: une cellule blindee en plomb de faible activite, un capteur INa, collimate, avec comptage direct dans une ''fenetre'' d'energie assez large recouvrant les pics des isotopes 239 et 241 du plutonium, dans la zone des 400 Kev, un analyseur programmable IN 96

Files

18002900.pdf

Files (292.4 kB)

Name Size Download all
md5:0f3f43c7b9ca5d04f6e4ede137d986cc
292.4 kB Preview Download

Additional details

Additional titles

Original title (French)
Station pour la mesure de faibles masses de dechets alpha

Publishing Information

Imprint Pagination
18 p.
Report number
CEA-CONF--8406

Conference

Title
Meeting on protection, handling, detection and safety (PMDS 86).
Dates
12-14 Mar 1986.
Place
Grenoble (France).