Intrinsic energy resolution of NaI(Tl)

Description

The light output for diameter 10 mmx10 mm and diameter 75 mmx75 mm NaI(Tl) crystals and energy resolution were measured for γ-ray energies ranging from 16-1333 keV. These measurements enabled the observation of the light yield nonproportionality behavior and allowed the determination of the intrinsic resolution after correcting for the measured resolution for photomultiplier tube (PMT) statistics. The intrinsic resolution was then compared with the nonproportionality component. The latter was calculated using measured electron response, Monte Carlo N Particle code (MCNP4B), and the simplified cascade sequence for NaI(Tl). This comparison allowed the identification of the intrinsic resolution component associated with δ-rays. Consequently, it was shown that the δ-ray component is the most dominant component of the NaI(Tl) intrinsic resolution

Additional details

Identifiers

PII
S0168900201019647;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
484
Journal Issue
1-3
Journal Page Range
p. 259-269
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.