Published October 1986 | Version v1
Miscellaneous

New conceptions for gamma spectrometry at very high counting rates

Description

The aim of the present work was not an improvement in one particular aspect of high-rate spectroscopy but the design of a whole system corresponding to the present state of technology. Special attention was given to the counting-loss correction, data evaluation and rejection of overlapping pulses. With the present system quantitative gamma spectroscopy up to 1 MHz input rate becomes possible. This system was integrated into a neutron activation analysis device. Parallel to this two other systems for low- and medium counting rates were constructed. Basing on the experience thus gained a multichannel analyzer with dead-time correction based on a commercial personal computer was developed. 21 refs., 47 figs., 23 tabs. (qui)

Availability note (English)

Available from the Technical University Vienna, Karlsplatz 13, A-1040 Vienna, Austria.

Additional details

Additional titles

Original title (German)
Neue Konzepte fuer die Gammaspektrometrie bei sehr hohen Zaehlraten

Publishing Information

Imprint Pagination