Published June 2013 | Version v1
Miscellaneous Restricted

Pulse shape analysis using CsI(Tl) Crystals

Description

The decay time of CsI(Tl) scintillating material consists of more than a single exponential component. The ratio between the intensity of these components varies as a function of the ionizing power of the absorbed particles, such as γ -rays or protons, and the temperature. This property can therefore be used for particle discrimination and for temperature monitoring, using pulse shape analysis. An unsupervised method that uses fuzzy clustering algorithms for particle identification based on pulse shape analysis is presented. The method is applied to discriminate between photon and proton-induced signals in CsI(Tl) scintillator detectors. The first results of a method that uses pulse shape analysis for correcting the temperature-dependent gain effect of the detector are also presented. The method aims at conserving a good energy resolution in a temperature varying environment without the need to measure the temperature of the detector externally (authors)

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
7 p.
Report number
ANIMMA--2013-1156

Conference

Title
3. international conference on Advancements in Nuclear Instrumentation Measurement Methods and their Applications
Acronym
ANIMMA 2013
Dates
23-27 Jun 2013
Place
Marseille (France)

Optional Information

Notes
12 Refs.; Available from the INIS Liaison Officer for France, see the 'INIS contacts' section of the INIS website for current contact and E-mail addresses: http://www.iaea.org/inis/Contacts/