Volumetric modulated arc therapy for delivery of hypofractionated stereotactic lung radiotherapy: A dosimetric and treatment efficiency analysis
Creators
- 1. Department of Radiation Oncology, William Beaumont Hospital, Royal Oak, MI (United States)
Description
Purpose/objective(s): Volumetric modulated arc therapy (VMAT) allows for intensity-modulated radiation delivery during gantry rotation with dynamic MLC motion, variable dose rates and gantry speed modulation. We compared VMAT plans with 3D-CRT for hypofractionated lung radiotherapy. Materials/methods: Twenty-one 3D-CRT plans for Stage IA lung cancer previously treated stereotactically were selected. VMAT plans were generated by optimizing machine aperture shape and radiation intensity at 10 deg. intervals. A partial arc range of 180 deg. was manually selected to coincide with tumor location. The arc was resampled down to 5 deg. intervals to ensure dose calculation accuracy. Identical planning objectives were used for VMAT/3D-CRT. Parameters assessed included dose to PTV and organs-at-risk (OAR), monitor units, and multiple conformity and homogeneity indices. Plans were delivered to a phantom for time comparison. Results: Lung V20/12.5/10/5 were less with VMAT (relative reduction 4.5%, p = .02; 3.2%, p = .01; 2.6%, p = .01; 4.2%, p = .03, respectively). Mean/maximum-doses to PTV, dose to additional OARs, 95% isodose line conformity, and target volume homogeneity were equivalent. VMAT improved conformity at both the 80% (1.87 vs. 1.93, p = .08) and 50% isodose lines (5.19 vs. 5.65, p = .01). Treatment times were reduced significantly with VMAT (mean 6.1 vs. 11.9 min, p < .01). Conclusions: Single arc VMAT planning achieves highly conformal dose distributions while controlling dose to critical structures, including significant reduction in lung dose volume parameters. Employing a VMAT technique decreases treatment times by 37-63%, reducing the chance of error introduced by intrafraction variation. The quality and efficiency of VMAT is ideally suited for stereotactic lung radiotherapy delivery.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radonc.2009.12.039Additional details
Identifiers
- DOI
- 10.1016/j.radonc.2009.12.039;
- PII
- S0167-8140(10)00016-2;
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 95
- Journal Issue
- 2
- Journal Page Range
- p. 153-157
- ISSN
- 0167-8140
- CODEN
- RAONDT
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41074439
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; DOSE RATES; EFFICIENCY; LUNGS; NEOPLASMS; OPTIMIZATION; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- BODY; DISEASES; DOSES; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RESPIRATORY SYSTEM; STRUCTURAL MODELS; THERAPY
Optional Information
- Copyright
- Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.