Published 1977 | Version v1
Journal article

Application of X-ray spectroscopy in nondestructive photon activation analysis

  • 1. Bundesanstalt fuer Materialpruefung, Berlin (Germany, F.R.)

Description

The use of X-ray spectroscopy for the qualitative and quantitative analysis of samples activated by 30 MeV bremsstrahlung from an electron linear accelerator. Detection limits are calculated from the measured X-ray spectra and compared with those for γ-ray spectroscopy. In general, the detection limits for γ-ray and X-ray spectroscopy are comparable. Higher sensitivities for X-ray spectroscopy are observed when only low intensity γ-rays are emitted by the activation products. X-ray spectroscopy should be applied in three cases: (a) low γ-ray emission probability, (b) extremely complicated γ-ray spectrum, (c) overlapping of γ-ray lines from different elements. γ-ray spectroscopy should be preferred for the analysis of light elements for two reasons: very strong absorption of low energy X-rays (low Z) within the sample, low X-ray emission probability for the activation products of light elements. Therefore no attempt was made to use X-ray spectroscopy for the analysis of elements below Ti. Some practical applications of X-ray spectroscopy in nondestructive multielement analysis are quoted. (T.G.)

Additional details

Publishing Information

Journal Title
J. Radioanal. Chem.
Journal Volume
37
Journal Issue
1
Series
J. Radioanal. Chem.
Journal Page Range
195-202

Conference

Title
International conference on modern trends in activation analysis.
Dates
13 Sep 1976.
Place
Munich, Germany, F.R.

Optional Information

Notes
4 figs.; 7 refs.; 2 tabs.