Radiofrequency mass spectrometers of L.G. Smith's type can reach a resolving power of 106-107 and a precision of 10-9-10-10. The resolving power, shape of peaks and limitations are described. As an example, the spectrometer to be used in an experiment aimed at measuring the anti p/p mass ratio is considered. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res., Sect. A
- Journal Volume
- 271
- Journal Issue
- 3
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 512-517
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 19106557
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; ANTIPROTONS; BEAM OPTICS; ELECTRIC FIELDS; FOCUSING; GEOMETRICAL ABERRATIONS; INHOMOGENEOUS FIELDS; MAGNETIC FIELDS; MASS DIFFERENCE; MASS RESOLUTION; MASS SPECTROMETERS; MASS SPECTROSCOPY; PROTONS; RELATIVISTIC RANGE; REST MASS; RF SYSTEMS
- Descriptors DEC
- ANTIBARYONS; ANTIMATTER; ANTINUCLEI; ANTINUCLEONS; ANTIPARTICLES; BARYONS; CATIONS; CHARGED PARTICLES; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; HADRONS; HYDROGEN IONS; HYDROGEN IONS 1 PLUS; IONS; MASS; MATTER; MEASURING INSTRUMENTS; NUCLEI; NUCLEONS; PARTICLE PROPERTIES; RESOLUTION; SPECTROMETERS; SPECTROSCOPY