Low limits and nonlinearity of clibration in suppressed ion chromatography with strong acid eluent
- 1. Dept. of Chemistry, Yonsei University, Seoul (Korea, Republic of)
- 2. University College, Yonsei University, Seoul (Korea, Republic of)
Description
Theoretical and experimental studies on nonlinearity of calibration in suppressed ion chromatography have been conducted. In suppressed cation chromatography with a strong acid as an eluent, nonlinear response in conductivity detection caused by water dissociation equilibrium shift has not been of concern in general. However, this issue can be of great consequence for analysis at the µg/L level. Theoretical calculation shows that the calibration curves at µg/L level are highly curvatured and a negative response range exists, making it impossible to make a determination. The extent of this phenomenon highly depends on the types and concentrations of ionic impurities present in the eluent. The presence of nonlinearity of calibration at the µg/L level is demonstrated experimentally. In order that linear calibrations are assured and negative response is eliminated, an addition of small amount of cation impurities into the eluent is recommended, as excessive amount tends to increase background conductance and their detection limits
Additional details
Publishing Information
- Journal Title
- Bulletin of the Korean Chemical Society
- Journal Volume
- 36
- Journal Issue
- 8
- Series
- 9 refs, 10 figs, 2 tabs
- Journal Page Range
- p. 2073-2080
- ISSN
- 0253-2964
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 48082126
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- CATIONS; CHROMATOGRAPHY; COMPUTER CALCULATIONS; DETECTION; EQUILIBRIUM; IMPURITIES
- Descriptors DEC
- CHARGED PARTICLES; IONS; SEPARATION PROCESSES