An inexpensive approach to improving data derived from single photon emission computed tomography

  • 1. Texas Univ., Dallas, TX (USA). Southwestern Medical School

Description

The issue of inadequate solid standards for correcting nonuniformity in single photon emission computed tomography was addressed. Controlled, complex, motion of commercially available cobalt-57 standards was found to enhance the perceived nonuniformity of the standard and to minimize the effects of any local flaws. This manoeuvre provides a better standard than a refillable liquid phantom without the problems inherent in radioactive liquid standards which include: phantom deformation, poor mixing, absence of retrospective verification, accidental contamination and substantial handling requirements. (author)

Additional details

Publishing Information

Journal Title
Nuclear Medicine and Biology
Journal Volume
15
Journal Issue
6
Series
Nucl. Med. Biol.
Journal Page Range
683-687
ISSN
0883-2897