Published March 2007 | Version v1
Journal article

Analyses of the Setup Errors using on Board Imager (OBI)

  • 1. Dept. of Radiation Oncology, Chonnam National University Hwasun Hospital, Hwasun (Korea, Republic of)

Description

The accuracy and advantages of OBI(On Board Imager) against the conventional method like film and EPID for the setup error correction were evaluated with the analysis of the accumulated data which were produced in the process of setup error correction using OBI. The results of setup error correction using OBI system were analyzed for the 130 patients who had been planned for 3 dimensional conformal radiation therapy during March 2006 and May 2006. Two kilo voltage images acquired in the orthogonal direction were fused and compared with reference setup images. The setup errors in the direction of vertical, lateral, longitudinal axis were recorded and calculated the distance from the isocenter. The corrected setup error were analyzed according to the lesion and the degree of shift variations. There was no setup error in the 41.5% of total analyzed patients and setup errors between 1 mm and 5 mm were found in the 52.3%. 6.1% patients showed the more than 5 mm shift and this error were verified as a difference of setup position and the movement of patient in a treatment room. The setup error analysis using OBI in this study verified that the conventional setup process in accordance with the laser and field light was not enough to get rid of the setup error. The KV images acquired using OBI provided good image quality for comparing with simulation images and much lower patients' exposure dose compared with conventional method of using EPID. These advantages of OBI system which were confirmed in this study proved the accuracy and priority of OBI system in the process of IGRT(Image Guided Radiation Therapy).

Additional details

Publishing Information

Journal Title
Journal of the Korean Society for Radiotherapeutic Technology
Journal Volume
19
Journal Issue
1
Series
7 refs, 5 figs, 1 tab
Journal Page Range
p. 1-5
ISSN
1598-8449

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
43125227
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; COMPUTERIZED SIMULATION; CORRECTIONS; ERRORS; IMAGES; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; SIMULATION; THERAPY