10∼100 kV free-air ionization chamber

  • 1. Academia Sinica, Beijing, BJ (China). Inst. of Atomic Energy

Description

The Free-Air Chamber for low energy X-rays and its correction factors and results are introduced. The total uncertainty of the exposure measurement of this chamber is estimated being 0.39% (K =1) for 25∼100 kV, and 0.45% for 10 kV X-rays, respectively. Calibration of the secondary standard dosimeter NPL-2560 produced by NE company, U.K. with this chamber shows a systematic difference lower than the results given by NPL in March, 1983, but little higher than that given by the China National Institute Metrology. However, they are still agreed within 1%. (12 refs., 6 tabs., 2 figs.)

Availability note (English)

MF available from INIS under the Report Number.

Files

28076314.pdf

Files (312.1 kB)

Name Size Download all
md5:28b0b5b04602c36fbfa3a80b1e6af0e3
312.1 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
11 p.
Report number
CNIC--01173

Optional Information

Secondary number(s)
IAE--0168.