Corrections to absorbed dose calculations for tissue inhomogeneities

  • 1. Ontario Cancer Institute, Toronto, Ontario M4X 1K9, Canada

Description

Traditional methods for correcting for the presence of tissue inhomogeneities may produce errors as great as 10% at points within or close to the inhomogeneity. A more accurate method is presented which employs tissue--air ratios raised to some power dependent on the relative electron densities of the inhomogeneities involved. Corrections may be made for points that lie within or below an inhomogeneity as well as for multiple inhomogeneities. Measurements were made in phantoms containing aluminum or cork inhomogeneities. Agreement between measured and predicted results was usually within 2%--3%

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
4
Journal Issue
5
Series
Med. Phys.
Journal Page Range
431-436
ISSN
0094-2405

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
9352352
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Descriptors DEI
CORRECTIONS; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; RADIOTHERAPY; TISSUE-EQUIVALENT MATERIALS; TISSUES
Descriptors DEC
BODY; MEDICINE; MOCKUP; STRUCTURAL MODELS; THERAPY