Published September 1977
| Version v1
Journal article
Corrections to absorbed dose calculations for tissue inhomogeneities
Creators
- 1. Ontario Cancer Institute, Toronto, Ontario M4X 1K9, Canada
Description
Traditional methods for correcting for the presence of tissue inhomogeneities may produce errors as great as 10% at points within or close to the inhomogeneity. A more accurate method is presented which employs tissue--air ratios raised to some power dependent on the relative electron densities of the inhomogeneities involved. Corrections may be made for points that lie within or below an inhomogeneity as well as for multiple inhomogeneities. Measurements were made in phantoms containing aluminum or cork inhomogeneities. Agreement between measured and predicted results was usually within 2%--3%
Additional details
Identifiers
- DOI
- 10.1118/1.594329;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 4
- Journal Issue
- 5
- Series
- Med. Phys.
- Journal Page Range
- 431-436
- ISSN
- 0094-2405
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 9352352
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- CORRECTIONS; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; RADIOTHERAPY; TISSUE-EQUIVALENT MATERIALS; TISSUES
- Descriptors DEC
- BODY; MEDICINE; MOCKUP; STRUCTURAL MODELS; THERAPY