Published July 1990 | Version v1
Journal article

Comparison of ferrous sulfate (Fricke) and ionization dosimetry for high-energy photon and electron beams

  • 1. Division of Radiation Physics, Cancer Control Agency of British Columbia, Vancouver, British Colombia (Canada)

Description

Dose measurements using Fricke and ionization methods were compared for 60 Co gamma rays, 4--25-MV photons, and 10--25-MeV electrons. Fricke derived doses based on a constant yield (εmG) were in good agreement with ionization derived doses based on the American Association of Physicists in Medicine Task Group 21 protocol and the National Research Council of Canada (NRC) ND calibration or the NRC proposed Nx. These measurements also confirmed the validity of the double-voltage technique in the collection efficiency correction, even for swept electron beams. Assuming the correctness of the ionization derived doses, the radiation yield appeared to be 1% higher and to increase with photon energy when irradiation vessels were made of Pyrex but not with polystyrene cells. These glass wall effects could be due to the scattering perturbation of electrons between inhomogeneous materials and, in particular for photon beams, due to the mismatch in mass energy absorption ratios and mass collision stopping power ratios between the Fricke dosimeter and the wall materials

Additional details

Publishing Information

Journal Title
Medical Physics
Journal Volume
17
Journal Issue
4
Series
Med. Phys.
Journal Page Range
602-606
ISSN
0094-2405
CODEN
MPHYA