Alternative positron emission tomography with non-conventional positron emitters: effects of their physical properties on image quality and potential clinical applications
Creators
- 1. Institute of Experimental Medicine, CNR, Rome (Italy)
- 2. Karolinska Pharmacy and Clinical Neuroscience, Section of Clinical Neurophysiology, Karolinska Hospital and Institute, Stockholm (Sweden)
- 3. Department of Hospital Physics, Section for Nuclear Medicine, Karolinska Hospital, Stockholm (Sweden)
Description
The increasing amount of clinically relevant information obtained by positron emission tomography (PET), primarily with fluorine-18 labelled 2-deoxy-2-fluoro-d-glucose, has generated a demand for new routes for the widespread and cost-efficient use of positron-emitting radiopharmaceuticals. New dual-head single-photon emission tomography (SPET) cameras are being developed which offer coincidence detection with camera heads lacking a collimator or SPET imaging with specially designed collimators and additional photon shielding. Thus, not only satellite PET imaging units but also nuclear medicine units investing in these new SPET/PET systems need to examine all available alternatives for rational radionuclide supplies from host cyclotrons. This article examines 25 ''alternative'' positron-emitting radionuclides, discusses the impact of their decay properties on image quality and reviews methods for their production as well as for their application in imaging techniques. (orig.)
Additional details
Publishing Information
- Journal Title
- European Journal of Nuclear Medicine
- Journal Volume
- 24
- Journal Issue
- 10
- Journal Page Range
- p. 1301-1327.
- ISSN
- 0340-6997
- CODEN
- EJNMD9
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 28074688
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BETA-PLUS DECAY RADIOISOTOPES; DIAGNOSTIC USES; IMAGES; PHYSICAL PROPERTIES; POSITRON COMPUTED TOMOGRAPHY; RADIOPHARMACEUTICALS; REVIEWS
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; COMPUTERIZED TOMOGRAPHY; DOCUMENT TYPES; DRUGS; EMISSION COMPUTED TOMOGRAPHY; ISOTOPES; LABELLED COMPOUNDS; MATERIALS; RADIOACTIVE MATERIALS; RADIOISOTOPES; TOMOGRAPHY; USES
Optional Information
- Notes
- With 4 tabs., 346 refs.