Published 1991 | Version v1
Journal article

Radiation protection metrology at a high-energy neutron therapy facility

  • 1. Clatterbridge Hospital, The Wirral (United Kingdom). MRC Cyclotron Unit
  • 2. Leeds General Infirmary (United Kingdom)
  • 3. National Radiological Protection Board, Chilton (United Kingdom)
  • 4. National Physical Lab., Teddington (United Kingdom)

Description

A radiation protection survey has been carried out at a high-energy neutron therapy facility using a combination of different detectors and counters. Included in the survey were measurements with a tissue equivalent proportional counter (TEPC), a rem meter, a large volume ionisation chamber (LVI) and a Geiger counter. Dose equivalent rates, normalised to a proton beam current of 25 μA, of between 1 μSv.h-1 and 0.7 Sv.h-1 were recorded depending on the location. In general the results confirm the tendency of the rem meter to over-read in fields consisting mainly of low energy neutrons and illustrate the advantages of the diagnostic and gamma discriminating properties of the TEPC. The LVI-Geiger system was found to be the least favourable combination of dosemeters, substantially under-reading and being unable to estimate the neutron dose rate at levels below about 32 μGy.h-1. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
37
Journal Issue
2
Series
Radiat. Prot. Dosim.
Journal Page Range
95-101
ISSN
0144-8420
CODEN
RPDOD