Enhanced 3D PET OSEM reconstruction using inter-update Metz filtering
Creators
- 1. MINERVA Optimization Center, Technion, Haifa 32000 (Israel)
- 2. MRC CU, Hammersmith Hospital, London (United Kingdom)
- 3. HSR, Milan (Italy)
- 4. Department of Mathematical Sciences, Brunel University, Uxbridge (United Kingdom)
Description
We present an enhancement of the OSEM (ordered set expectation maximization) algorithm for 3D PET reconstruction, which we call the inter-update Metz filtered OSEM (IMF-OSEM). The IMF-OSEM algorithm incorporates filtering action into the image updating process in order to improve the quality of the reconstruction. With this technique, the multiplicative correction image - ordinarily used to update image estimates in plain OSEM - is applied to a Metz-filtered version of the image estimate at certain intervals. In addition, we present a software implementation that employs several high-speed features to accelerate reconstruction. These features include, firstly, forward and back projection functions which make full use of symmetry as well as a fast incremental computation technique. Secondly, the software has the capability of running in parallel mode on several processors. The parallelization approach employed yields a significant speed-up, which is nearly independent of the amount of data. Together, these features lead to reasonable reconstruction times even when using large image arrays and non-axially compressed projection data. The performance of IMF-OSEM was tested on phantom data acquired on the GE Advance scanner. Our results demonstrate that an appropriate choice of Metz filter parameters can improve the contrast-noise balance of certain regions of interest relative to both plain and post-filtered OSEM, and to the GE commercial reprojection algorithm software. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology (Online)
- Journal Volume
- 45
- Journal Issue
- 8
- Journal Page Range
- p. 2417-2439
- ISSN
- 1361-6560
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43066963
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; COMPUTER CODES; IMAGE PROCESSING; POSITRON COMPUTED TOMOGRAPHY; QUALITY ASSURANCE
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; MATHEMATICAL LOGIC; PROCESSING; TOMOGRAPHY
Optional Information
- Contract/Grant/Project number
- Grant EP23493
- Notes
- Refs; This record replaces 31040294