Dosimetric comparison of two 90Sr/90Y sources for intravascular brachytherapy: an EGSnrc Monte Carlo calculation
Creators
- 1. Department of Radiation Oncology, University of Maryland, Baltimore, MD (United States)
Description
A new 90Sr/90Y catheter-based system, the Beta-CathTM 3.5F system, was recently introduced into the clinic for intravascular brachytherapy. This new system has a smaller diameter of delivery catheter (3.5 French) compared to the original Beta-CathTM system (5 French catheter). The new source/seed (3.5F source) has a different design from the original (5F) source. In this paper, we report the EGSnrc Monte Carlo results of two-dimensional dose rate distributions in water as well as the TG-60 dose parameters for the new 3.5F source and the original 5F source. A complete set of the dosimetric data for each individual seed is tabulated. A detailed dosimetric comparison was carried out for the two sources. The comparison shows that differences between the two sources are less than 5% in the region of interest for intravascular brachytherapy. Higher differences (up to 10%) are seen in the near-source region, which is mainly inside the delivery catheter. Based on the present calculation and comparison, we conclude that the new Beta-CathTM 3.5F system is dosimetrically equivalent to the original Beta-CathTM (5F) system. Therefore, clinical dosimetry and the practice of using both the systems can be treated similarly
Availability note (English)
Available online at http://stacks.iop.org/0031-9155/47/4259/m22311.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://stacks.iop.org/0031-9155/47/4259/m22311.pdf; http://www.iop.org/;
- PII
- S0031-9155(02)39501-0;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 47
- Journal Issue
- 23
- Journal Page Range
- p. 4259-4269
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34021540
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; DOSIMETRY; MONTE CARLO METHOD; RADIATION DOSES; RADIATION SOURCE IMPLANTS; RADIOTHERAPY; STRONTIUM 90; YTTRIUM 90
- Descriptors DEC
- ALKALINE EARTH ISOTOPES; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CALCULATION METHODS; DAYS LIVING RADIOISOTOPES; DOSES; EVALUATION; EVEN-EVEN NUCLEI; HOURS LIVING RADIOISOTOPES; IMPLANTS; INTERMEDIATE MASS NUCLEI; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEDICINE; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; RADIATION SOURCES; RADIOISOTOPES; RADIOLOGY; STRONTIUM ISOTOPES; THERAPY; YEARS LIVING RADIOISOTOPES; YTTRIUM ISOTOPES