Published 1990
| Version v1
Report
Determination of uranium in urine by fluorometry
Description
The method to measure natural uranium in urine used in the CIEMAT has been revised, evaluated, and documented. The procedure is based on a direct fluorometric determination in the ashed urine sample. Two factors wich could influence the results have been studied: Ca(II) amounts in the sample, and inorganic waste in fused pellet. The lower limit of detection has been determined from 'blank' urine samples and it has been found in agreement with 'ICRP' recommendations. The method has been shown to be quench independent, rapid, accurate and economical. (Author)
Availability note (English)
MF available from INIS under the Report Number.Additional details
Additional titles
- Original title (Spanish)
- Determinacion de uranio en orina por fluorimetria
Publishing Information
- Imprint Pagination
- 25 p.
- Report number
- CIEMAT--654
INIS
- Country of Publication
- Spain
- Country of Input or Organization
- Spain
- INIS RN
- 22021854
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- FLUORESCENCE; NATURAL URANIUM; PERSONNEL MONITORING; QUANTITATIVE CHEMICAL ANALYSIS; URINE
- Descriptors DEC
- ACTINIDES; BIOLOGICAL MATERIALS; BIOLOGICAL WASTES; BODY FLUIDS; CHEMICAL ANALYSIS; ELEMENTS; EMISSION; LUMINESCENCE; MATERIALS; METALS; MONITORING; PHOTON EMISSION; RADIATION MONITORING; URANIUM; WASTES