Published February 1988 | Version v1
Journal article

Extension of measurable range of film dosimetry by a low energy X-ray absorption method

  • 1. Toyohashi Univ. of Technology, Aichi (Japan). Dept. of Electrican and Electronic Engineering

Description

The availability of an X-ray absorption technique employing a very low power X-ray tube was examined to extend measurable range for photographic film dosimetry. The X-ray tube having titanium target was operated at 8 kV and 0.2 μA to emit Ti KX-rays of moderate intensity. The degree of the Ti KX-ray absorption, defined as similar to photographic density, was measured for the two kinds of badge film, Fuji gamma-ray badge film and Kodak personal monitoring film, type 2 exposed for 60Co or 137Cs gamma-rays and developed by the respective standard procedures. The experimental results show that the dosimetric range of 0.01-100 R for the Fuji film and 0.03-1000 R for the Kodak film may be easily measured by 1 minute counting with the relative statistical error (σ) of 10%. (author)

Additional details

Publishing Information

Journal Title
Radioisotopes (Tokyo)
Journal Volume
37
Journal Issue
2
Series
Radioisotopes (Tokyo).
Journal Page Range
88-91
ISSN
0033-8303
CODEN
RAISA