Published December 1, 1985 | Version v1
Journal article

Stable Josephson reference voltages between 0.1 and 1.3 V for high-precision voltage standards

  • 1. Physikalisch-Technische Bundesanstalt, Bundesallee 100, 3300 Braunschweig, Federal Republic of Germany

Description

A new series array of 1440 Josephson tunnel junctions has been developed and tested as a reference voltage standard. It yields microwave induced quantized voltage steps up to 1.3 V. The steps are usually stable for more than 5 h with a microwave driving frequency of either 70 or 90 GHz. A high-resolution comparison of a constant voltage step at the 1-V level with the electromotive force of a saturated Weston cell is described. The comparison shows that the step voltage is constant to within +- 1 nV over the full step width

Additional details

Publishing Information

Journal Title
Appl. Phys. Lett.
Journal Volume
47
Journal Issue
11
Series
Appl. Phys. Lett.
Journal Page Range
1222-1223
ISSN
0003-6951
CODEN
APPLA