Published November 1983 | Version v1
Journal article

Dual photon absorptiometry of the spine with a low activity source of gadolinium 153

  • 1. Royal Infirmary, Edinburgh (UK)

Description

Apparatus and data-processing techniques were developed to measure the bone mineral content (BMC) of the lumbar spine. 153Gd was used as the dual photon source with an activity down to 7 GBq, lower than that adopted by other workers. The compromise between resolution and sensitivity was optimised. A reproducibility in normal subjects of 1.5% (coefficient of variation) was obtained. Normalisation procedures using parameters of the scan image and other indicators of body size such as span were developed, and reduced the coefficient of variation between normal subjects to 10% for men and 8% for pre-menopausal women. The absolute values of BMC are similar to those reported by others. Cross-sectional studies in women demonstrated an accelerated loss of BMC after the menopause. (author)

Additional details

Publishing Information

Journal Title
Br. J. Radiol.
Journal Volume
56
Journal Issue
671
Series
Br. J. Radiol.
Journal Page Range
829-835
ISSN
0007-1285