Published October 2007 | Version v1
Journal article

Early experiences with the Mount Allison University four-detector X-ray fluorescence bone lead measurement system

  • 1. Department of Physics, Mount Allison University, 67 York Street, Sackville, New Brunswick, E4L 1E6 (Canada)

Description

A four-detector (4 x 500 mm2) low energy Germanium detector array has recently been acquired for use in a 109Cd source K-shell X-ray fluorescence bone lead measurement system. In this paper, system calibration results are presented, along with preliminary measurement results using lead-doped bone phantoms. Measurement uncertainty and reproducibility values are noted. System comparisons are made with our conventional (1 x 2000 mm2) detector system, and future directions outlined

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nimb.2007.04.133

Additional details

Identifiers

DOI
10.1016/j.nimb.2007.04.133;
PII
S0168-583X(07)00791-4;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
Journal Volume
263
Journal Issue
1
Journal Page Range
p. 32-35
ISSN
0168-583X
CODEN
NIMBEU

Conference

Title
6. topical meeting on industrial radiation and radioisotope measurement applications
Dates
20-24 Jun 2005
Place
Hamilton, ON (Canada)

Optional Information