Published February 15, 1977
| Version v1
Journal article
Depth dose distribution of 9 MeV photons in a single phosphate glass compared to phantom results
Creators
- 1. Gesellschaft fuer Kernforschung m.b.H., Karlsruhe (Germany, F.R.)
Description
Phosphate glasses, LiF:Mg, Ti detectors as well as ionization chambers have been applied to dose measurements in a 9 MeV photon beam. Agreement was found between the results of phantom measurements and the depth dose distribution within a single block of glass. Irradiated free in air these detectors show a response between 0.27 and 1.28 in comparison to the reading of an ionization chamber at a depth of 3 g/cm2. (Auth.)
Additional details
Identifiers
Publishing Information
- Journal Title
- Nuclear Instruments and Methods
- Journal Volume
- 141
- Journal Issue
- 1
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 141-144
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 8325388
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DEPTH DOSE DISTRIBUTIONS; IONIZATION CHAMBERS; IRRADIATION; MEV RANGE 01-10; PHANTOMS; PHOSPHATE GLASS; PHOTONS; RESPONSE FUNCTIONS; RPL DOSEMETERS; THERMOLUMINESCENT DOSEMETERS; TISSUE-EQUIVALENT MATERIALS
- Descriptors DEC
- DOSEMETERS; ELEMENTARY PARTICLES; ENERGY RANGE; FUNCTIONS; LUMINESCENT DOSEMETERS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MEV RANGE; MOCKUP; RADIATION DETECTORS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS