Published March 2000 | Version v1
Journal article

Monitor unit calculation for an intensity modulated photon field by a simple scatter-summation algorithm

  • 1. Department of Radiation Oncology, Stanford University, Stanford, CA 94305-5304 (United States).
  • 2. Department of Radiation Oncology, Stanford University, Stanford, CA 94305-5304 (United States)

Description

An important issue in intensity modulated radiation therapy (Imr) is the verification of the monitor unit (MU) calculation of the planning system using an independent procedure. Because of the intensity modulation and the dynamic nature of the delivery process, the problem becomes much more involved than that in conventional radiation therapy. In this work, a closed formula for MU calculation is derived. The approach is independent of the specific form of leaf sequence algorithms. It is straightforward to implement the procedure using a simple computer program. The approach is illustrated by a simplified example and is demonstrated by a few CORVUS (NOMOS Corporation, Sewickley, PA) treatment plans. The results indicate that it is robust and suitable for IMRT MU verification. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology (Online)
Journal Volume
45
Journal Issue
3
Journal Page Range
p. N1-N7
ISSN
1361-6560

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
31022260
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ALGORITHMS; BEAM MONITORING; CALCULATION METHODS; IRRADIATION DEVICES; IRRADIATION PROCEDURES; MODIFICATIONS; MODULATION; OPTIMIZATION; RADIATION MONITORING; RADIOTHERAPY
Descriptors DEC
MATHEMATICAL LOGIC; MEDICINE; MONITORING; THERAPY