Published 2011 | Version v1
Journal article

Use of the HRICP-MS technique for the evaluation of boron isotopes in Eucalitpus plants

  • 1. Universidade Federal de Vicosa (UFV), MG (Brazil). Dept. de Solos
  • 2. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

Description

The 10B isotope tracer technique is essential to study the B mobility in plants. Factors that can influence the quality of measured B isotope ratios were optimized experimentally using High Resolution Inductively Coupled Plasma Mass Spectrometry (HRICP-MS). An isotopically certified standard (NIST SRM-951) was used. The best combination was obtained using a resolution of 400, a RF power of 1250 W, followed by 15 measurements over a 10-s integration period each (15*10). Utilizing this approach it was possible to obtain a precision of 0.3 % in standard material and 2z % in the experimental samples. The results show the importance of establishing optimized work conditions before carrying out the analytical series. (author)

Availability note (English)

Available from http://www.scielo.br/pdf/qn/v34n3/25.pdf

Additional details

Additional titles

Original title (Portuguese)
Uso da tecnica HRICP-MS na avaliacao dos isotopos de boro em eucalipto

Publishing Information

Journal Title
Quimica Nova On-Line
Journal Volume
34
Journal Issue
3
Journal Page Range
p. 512-515
ISSN