Published July 11, 2000 | Version v1
Journal article

Experiments with a radioactive 56Ni beam

Description

A technique for producing a high-quality radioactive 56Ni (T1/2=6.1 d) beam via the two-accelerator method has been developed. Beam intensities of 2x10756Ni/s were extracted from the ion source and 2x10456Ni/s were delivered to the target. For a study of neutron transfer reactions in inverse kinematics, a high-efficiency detection system was built consisting of a large solid angle (2.8 sr), high-granularity Si detector array for measuring the outgoing protons in coincidence with the heavy reaction products identified with respect to mass A and nuclear charge Z in the focal plane of a recoil mass separator

Additional details

Identifiers

PII
S0168900299014394;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
449
Journal Issue
1-2
Journal Page Range
p. 208-216
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2000 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.