Published October 2018 | Version v1
Journal article

Detection of long-lived contaminants in cyclotron-produced radiopharmaceuticals by large area plastic scintillators

  • 1. RSE, Inc., Chandler, AZ (United States)
  • 2. White Rock Science, White Rock, NM (United States)
  • 3. Eastern Scientific, Oak Ridge, TN (United States)
  • 4. Oak Ridge National Laboratory, Oak Ridge, TN (United States)

Description

In the USA during 2015 over 1,718,500 clinical PET scans were performed in 2380 centers. Patients undergoing these studies will alarm homeland security monitors at borders, ports, and some airports for up to 2 days due to the positron-annihilation photons (0.511 MeV) from the imaging isotope 18F (110 min half-life). In some radiopharmaceuticals, long-lived contaminants generated from the activation and spallation of elements in the HAVAR window of the cyclotron target may also be present in the patient dose. In this work, we have modeled a typical homeland security portal monitor in MCNP, and determined its sensitivity to each of the possible contaminating isotopes found in PET radiopharmaceuticals when distributed in patient phantoms. These data were combined with the time-dependent uptake/retention functions for the contaminating isotopes, to identify the isotopes that can cause alarms for substantial periods of time after the short-lived imaging agent has decayed away. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
318
Journal Issue
1
Journal Page Range
p. 11-15
ISSN
0236-5731
CODEN
JRNCDM

Conference

Title
11. International Conference on Methods and Applications of Radioanalytical Chemistry
Acronym
MARC XI
Dates
8-13 Apr 2018
Place
Kailua-Kona (United States)

Optional Information

Notes
7 refs.