Quality control of micro-computed tomography systems

  • 1. Dept. of Radiology, UZ Brussel, Laarbeeklaan 101, 1090 Brussels (Belgium)
  • 2. Dept. of Radiation Protection, UZ Brussel, Laarbeeklaan 101, 1090 Brussels (Belgium)
  • 3. Dept. of Nuclear Medicine, UZ Brussel, Laarbeeklaan 101, 1090 Brussels (Belgium)

Description

The rapid proliferation of micro-computed tomography (micro-CT) scanners in preclinical small animal studies has created a need for a method on scanner performance evaluation and scan parameter optimisation. The purpose of this study was to investigate the performance of the scanner with a dedicated micro-CT phantom. The phantom was developed with different independent sections that allow for measurement of major scanner characteristics such as uniformity, linearity, contrast response, dosimetry and resolution. The results of a thorough investigation are discussed. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncq088

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
139
Journal Issue
1-3
Journal Page Range
p. 463-467
ISSN
0144-8420

Conference

Title
3. Malmoe Conference on Medical Imaging - Optimisation in X-ray and molecular imaging
Dates
25-27 Jun 2009
Place
Malmoe (Sweden)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
France
INIS RN
43102278
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; DOSIMETRY; EVALUATION; PHANTOMS; QUALITY ASSURANCE; QUALITY CONTROL; SENSITIVITY; SPECIFICITY; X RADIATION
Descriptors DEC
CONTROL; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MOCKUP; RADIATIONS; STRUCTURAL MODELS; TOMOGRAPHY

Optional Information

Notes
13 refs