Published 2010
| Version v1
Journal article
Quality control of micro-computed tomography systems
- 1. Dept. of Radiology, UZ Brussel, Laarbeeklaan 101, 1090 Brussels (Belgium)
- 2. Dept. of Radiation Protection, UZ Brussel, Laarbeeklaan 101, 1090 Brussels (Belgium)
- 3. Dept. of Nuclear Medicine, UZ Brussel, Laarbeeklaan 101, 1090 Brussels (Belgium)
Description
The rapid proliferation of micro-computed tomography (micro-CT) scanners in preclinical small animal studies has created a need for a method on scanner performance evaluation and scan parameter optimisation. The purpose of this study was to investigate the performance of the scanner with a dedicated micro-CT phantom. The phantom was developed with different independent sections that allow for measurement of major scanner characteristics such as uniformity, linearity, contrast response, dosimetry and resolution. The results of a thorough investigation are discussed. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncq088Additional details
Identifiers
- DOI
- 10.1093/rpd/ncq088;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 139
- Journal Issue
- 1-3
- Journal Page Range
- p. 463-467
- ISSN
- 0144-8420
Conference
- Title
- 3. Malmoe Conference on Medical Imaging - Optimisation in X-ray and molecular imaging
- Dates
- 25-27 Jun 2009
- Place
- Malmoe (Sweden)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 43102278
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; DOSIMETRY; EVALUATION; PHANTOMS; QUALITY ASSURANCE; QUALITY CONTROL; SENSITIVITY; SPECIFICITY; X RADIATION
- Descriptors DEC
- CONTROL; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MOCKUP; RADIATIONS; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- 13 refs