Published February 1987 | Version v1
Journal article

Wide-range neutron counting for yield prediction of PET agents in cyclotron irradiations

  • 1. Dept. of Medical Physics, Univ. of Wisconsin, Madison, WI 53706

Description

As PET is moving into the clinical setting, the instrumentation from yesterday's research laboratory is being pressed into routine use in a setting which is intolerant of any unreliability. At the University of Wisconsin, a system of three neutron detectors has been developed to assess the progress of target irradiations in real time. The triad of detectors consists of an NE213 liquid scintillator, a /sup 238/U fission detector, and a proton recoil telescope to be able to detect neutron fluence rates over a span of six decades. The signals from these detectors are the inputs to a leaky integrator which calculates the buildup of activity in the target. In this way, the cyclotron operator can easily monitor the irradiation and take immediate action, if required, as opposed to finding an empty target at the end of bombardment and falling hopelessly behind schedule. The ability to accurately predict target yields by wide-range neutron counting adds a measure of reliability to radiopharmaceutical production and will facilitate PET becoming a routine clinical tool

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-34
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
277-279
ISSN
0018-9499
CODEN
IETNA

Conference

Title
Nuclear science and nuclear power systems symposium.
Dates
29-31 Oct 1986.
Place
Washington, DC (USA).

Optional Information

Secondary number(s)
CONF-861007--.