Comparison of radiation doses of different machines in standard coronary angiography
Creators
- 1. Shandong Provincial Third Hospital, Image Center, Jinan (China)
- 2. Shandong Provincial Third Hospital Interventional Medicine, Jinan (China)
Description
Objective: To discuss the radiation dose contrast between heart special FD DSA machine and FD DSA machine in the standard 8 position coronary artery angiography. Methods: 50 cases were rearded with siemens AXIOM-Artisdfc heart FD DSA machine while the same samples with GE Tnnova IGS530 FD DSA machine. Todo stardard 8 position coronary artery angiography was done with 15 cm thick organic vireous die line measurement card for the two machines again. And then normal perspective and movie collection was done with 20 cm thick organic vitreous die line measurement card for the two machinesin referfnce point. At last the data was analyzed. Results: The value of cumulative dose for simens and GE were 153.7 mGy and 410 mGy respectively. The value of dpse-area product for simens and GE were 11.42 mGy * cm2 and 37.22 mGy * cm2 respectively. Conclusion: The radiation dose of the heart FD DSA machine was obviously less than that of the FD DSA machine. (authors)
Additional details
Publishing Information
- Journal Title
- Chinese Journal of Radiological Health
- Journal Volume
- 26
- Journal Issue
- 5
- Journal Page Range
- p. 550-551, 574
- ISSN
- 1004-714X
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 50069885
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; COMPARATIVE EVALUATIONS; CORONARIES; HEART; PHANTOMS; RADIATION DOSES
- Descriptors DEC
- ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DOSES; EVALUATION; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; STRUCTURAL MODELS
Optional Information
- Notes
- 6 tabs., 6 refs.