Published October 2018 | Version v1
Miscellaneous

Experimental characterization on accuracy of multi-detector boron meter

  • 1. Ulsan National Institute of Science and Technology, Ulsan (Korea, Republic of)
  • 2. USERS Incorporated Company, Daejeon (Korea, Republic of)

Description

The previous MCNP simulation studies show that the accuracy of the multi detector boron meter is improved (0.4%) and the accuracy improvement is 65% compared with the single detector boron meter through experiments which based on simulation. The accuracy of the multi - detector boron meter was confirmed through previous studies and the system was constructed for the experiment in the actual cooling water circulation environment. In this paper, a study was conducted to measure and analyze the accuracy of 0 to 5000 ppm boric acid concentration measurement of a multi-detector boron meter in a real reactor environment through a coolant circulation system. Accuracy analysis of multiple detector boron meter in actual coolant circulation environment was performed and the results were analyzed Compared to conventional single-detector boron meters, 43.45% improved accuracy is achieved. For the repeatability test, the re-measured count rate was used to evaluate the accuracy and the performance was improved by 12%.

Part of:
Proceedings of the Conference and Symposium Korean Radioactive Waste Society Autumn Meeting 2018

Additional details

Publishing Information

Publisher
KRS
Imprint Place
Daejeon (Korea, Republic of)
Imprint Title
Proceedings of the Conference and Symposium Korean Radioactive Waste Society Autumn Meeting 2018
Imprint Pagination
616 p.
Journal Page Range
p. 517-518

Conference

Title
2018 Autumn Meeting of Korean Radioactive Waste Society
Dates
31 Oct - 2 Nov 2018
Place
Daejeon (Korea, Republic of)

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
50081751
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ACCURACY; BORIC ACID; DETECTION; M CODES; PERFORMANCE; REACTORS; SIMULATION
Descriptors DEC
BORON COMPOUNDS; COMPUTER CODES; HYDROGEN COMPOUNDS; INORGANIC ACIDS; INORGANIC COMPOUNDS; OXYGEN COMPOUNDS

Optional Information

Notes
2 refs, 3 figs, 1 tab