Novel detectors for silicon based microdosimetry, their concepts and applications
Creators
Description
This paper presents an overview of the development of semiconductor microdosimetry and the most current (state-of-the-art) Silicon on Insulator (SOI) detectors for microdosimetry based mainly on research and development carried out at the Centre for Medical Radiation Physics (CMRP) at the University of Wollongong with collaborators over the last 18 years. In this paper every generation of CMRP SOI microdosimeters, including their fabrication, design, and electrical and charge collection characterisation are presented. A study of SOI microdosimeters in various radiation fields has demonstrated that under appropriate geometrical scaling, the response of SOI detectors with the well-known geometry of microscopically sensitive volumes will record the energy deposition spectra representative of tissue cells of an equivalent shape. This development of SOI detectors for microdosimetry with increased complexity has improved the definition of microscopic sensitive volume (SV), which is modelling the deposition of ionising energy in a biological cell, that are led from planar to 3D SOI detectors with an array of segmented microscopic 3D SVs. The monolithic ΔE−E silicon telescope, which is an alternative to the SOI silicon microdosimeter, is presented, and as an example, applications of SOI detectors and ΔE−E monolithic telescope for microdosimetery in proton therapy field and equivalent neutron dose measurements out of field are also presented. An SOI microdosimeter "bridge" with 3D SVs can derive the relative biological effectiveness (RBE) in 12C ion radiation therapy that matches the tissue equivalent proportional counter (TEPC) quite well, but with outstanding spatial resolution. The use of SOI technology in experimental microdosimetry offers simplicity (no gas system or HV supply), high spatial resolution, low cost, high count rates, and the possibility of integrating the system onto a single device with other types of detectors.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2015.08.059Additional details
Identifiers
- DOI
- 10.1016/j.nima.2015.08.059;
- PII
- S0168-9002(15)01023-2;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 809
- Journal Page Range
- p. 156-170
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48005947
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANIMAL TISSUES; CHARGE COLLECTION; COUNTING RATES; ENERGY ABSORPTION; ENERGY LOSSES; HEAVY IONS; ION BEAMS; MICRODOSIMETRY; PLANT TISSUES; PROPORTIONAL COUNTERS; PROTON BEAMS; RADIATION DOSES; RADIOTHERAPY; RBE; SIMULATION; SPATIAL RESOLUTION
- Descriptors DEC
- ABSORPTION; BEAMS; BODY; CHARGED PARTICLES; DOSES; DOSIMETRY; IONS; LOSSES; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; NUCLEON BEAMS; PARTICLE BEAMS; RADIATION DETECTORS; RADIOLOGY; RESOLUTION; SORPTION; THERAPY
Optional Information
- Copyright
- Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.