Published April 2008 | Version v1
Journal article

Development of multiple prompt gamma-ray analysis

  • 1. Japan Atomic Energy Agency, Nuclear Science and Energy Directorate, Tokai, Naka, Ibaraki (Japan)

Description

In order to improve the limits of quantification for trace elements, we applied the multiple gamma-ray detection method in the prompt gamma-ray analysis (PGA). The Clover Ge detector system was constructed for multiple prompt gamma-ray analysis (MPGA) at the neutron guide hall of the JRR-3M reactor of the Japan Atomic Energy Agency (JAEA). The MPGA system consists of three Clover detectors and a BGO Compton suppressor. The instrumentation of the MPGA beam line and the detector system were developed. The estimated limit of quantification of cadmium in the plastic sample is approximately 0.1 ppm. The capability of the MPGA system could be improved by modification of the detector system, shield arrangement, software algorithm, etc. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
276
Journal Issue
1
Journal Page Range
p. 217-220
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
10 refs.